- Take samples from
24 bubble free cells at up to 20 sampling time points
- Open air system,
no pressure build up during the tests, no leakage
- Syringe pump
sampling system accuracy to ± 1 %
- Sampling takes
place before media replacement to avoid sample dilution
- Printer for full report
The LOGAN SYSTEM 912-24 is designed to save valuable time by delivering
precise transdermal samples into HPLC vials automatically. The membrane is
mounted between the cell cap (donor) and the cell body (receptor). The membrane
is bathed from below by an isotonic saline solution. Temperature is maintained
at 37°C by thermostatically controlled water which enters the lower port of the
water jacket surrounding the saline solution chamber and circulates out through
the upper port. Water is supplied and removed by two (upper and lower)
stainless steel manifolds connected to a constant temperature bath.
Homogenous distribution of temperature in the saline bathing solution is
accomplished by the agitating motion of a Teflon-covered magnetic stirring bar,
driven by an external magnet mounted on a timing motor. The cell cap is open to
the air, exposing the membrane’s upper surface to ambient conditions. The open
cap also allows for a finite dosage application of the compound being studied.
The SYP Series Syringe Pump Module withdraws and measures sample volumes
from transdermal diffusion cells at each time interval. Individual samples are
dispensed simultaneously into HPLC vials that are stored in the SCT Series
Sample Collectors. To avoid sample dilution, the replacement of media takes
place right after each sampling. The lift plate tilts all Franz cells to allow
any air bubbles to purge out of the cells while doing the media replacement.
This will insure that there will be no air bubbles under the skin/patch; the
tilted plate goes back to the horizontal position once the media replacement is
complete and continues the test.